MARV Analyzer
The MARV Analyzer is a flow- and temperature-controlled system for continuous measurement of dust, moisture, and gas composition in industrial processes. It operates reliably in dry or wet conditions, including explosive atmospheres (ATEX Zone 1/2), making it ideal for steel production, emission control, and other harsh industrial environments.
At its core is a 3rd generation five-zone optical cell. The central zone holds undiluted sample gas, while surrounding purge zones protect optical windows from dust, moisture, and corrosive condensates. This design prevents fouling, maintains accurate measurements, and allows operation at up to 200 °C and 10 bar. Optical windows are made of durable infrared- and visible-transparent materials for chemical and thermal stability.
Dual-path optics enable simultaneous measurement of gas and dust. Infrared absorption analyzes gas and moisture, while visible-light forward scattering measures dust concentrations from 0.1 to 300 mg/m³. Sample gas is measured undiluted, with algorithmic compensation ensuring accurate multi-component analysis.
The modular design supports NDIR, TDLA, and FTIR spectroscopy, allowing flexible configuration and wide dynamic range measurement. The MARV Analyzer measures gases from ppm levels to high concentrations (e.g., CO up to 70% vol) while monitoring dust and moisture, providing a robust and precise solution for industrial process monitoring.
Key Features
- Simultaneous measurement of dust, moisture, and gas composition
- Dual-path optical system: IR absorption + visible light scattering
- Five-zone optical cell with purge protection against contamination
- Accurate measurement in undiluted gas streams
- Supports NDIR, TDLA, and FTIR spectroscopy methods
- Modular IR sources and detectors for flexible configuration
- Wide dynamic measurement range (ppm to high % concentrations)
- Operates in explosive atmospheres (ATEX Zone 1/2)
- Resistant to high temperature (up to 200 °C) and pressure (up to 10 bar)
- No sample conditioning required for dust measurement
Applications
BOF (Basic Oxygen Furnace)
Direct measurement of CO, dust, and moisture in converter off-gas without dilution, enabling process optimization and emission control.
- Temperature: Up to 200°C
- Pressure: Up to 10 bar
- Typical Gas Composition: [object Object]
Industrial gas monitoring (dry and wet processes)
Continuous monitoring of gas composition, dust, and moisture in industrial plants such as agglomeration, graphite production, and emission control systems.
- Temperature: Up to 200°C
- Humidity: Up to 40%
- Gas Flow: 5–40 l/min